All manuals and user guides at all-guides.com
Using the Store Function
Store/Recall button:
The screen will be flashing the MEM symbol and the number of the
stored data twice when the Store/Recall button is pressed.
1. Press once to store the voltage value during voltage test
2. Press once to store the PI/ DAR, insulation or Earth-bond
resistance value when the individual test is completed.
3. The memory is divided into five segments, each segment has
100 maximum data points. The method to store the data is first
in/first out when the memory is full.
Table: The stored values of the individual test
segments
Voltage
items
1
Voltage
2
3
Note: Press the Store/Recall button
The screen will be flashing MEM symbol and "cLr" on reading twice.
Using the Recall Function
Press the Store/Recall button
If the memory is empty, the meter will display the " nOnE " symbol.
Search the stored value under RECALL mode
Insulation
Earth-Bond
resistance
resistance
50~1000V
Resistance
Resistance
Leakage
Current
Test voltage
>
5 sec to clear the stored data.
>
2 sec to enter the RECALL mode.
10
DAR
PI
DAR value
PI value
R30-sec
R1-min
R1-min
R10-min