Error Messages And Troubleshooting - Omron M2 Basic Manual De Instrucciones

Ocultar thumbs Ver también para M2 Basic:
Tabla de contenido

Publicidad

3. Error Messages and Troubleshooting

If any of the below problems occur during measurement, check to make sure that no other electrical device is within 30 cm. If the problem persists,
please refer to the table below.
Display/Problem
appears or the
arm cuff does not
inflate.
appears or a
measurement
cannot be
completed after the
arm cuff inflates.
appears
appears
appears
/
appears
does not flash during
a measurement
All manuals and user guides at all-guides.com
Possible Cause
The [START/STOP] button was pressed
while the arm cuff is not applied.
Air plug is not completely plugged
into the monitor.
The arm cuff is not applied correctly.
Air is leaking from the arm cuff.
You move or talk during a
measurement and the arm cuff does
not inflate sufficiently.
Due to the systolic pressure is above
210 mmHg, a measurement cannot
be taken.
The arm cuff is inflated exceeding the
maximum allowable pressure.
You move or talk during a
measurement. Vibrations disrupt a
measurement.
The pulse rate is not detected
correctly.
Solution
Press the [START/STOP] button again to turn the monitor
off. After inserting the air plug securely and applying the
arm cuff correctly, press the [START/STOP] button.
Insert the air plug securely.
Apply the arm cuff correctly, then take another
measurement. Refer to section 4 of instruction manual
Replace the arm cuff to the new one. Refer to section 9 of
instruction manual
.
Remain still and do not talk during a measurement. If
"E2" appears repeatedly, inflate the arm cuff manually
until the systolic pressure is 30 to 40 mmHg above your
previous readings. Refer to section 6 of instruction
manual
.
Do not touch the arm cuff and/or bend the air tube while
taking a measurement. If inflating the arm cuff manually,
refer to section 6 of instruction manual
Remain still and do not talk during a measurement.
Apply the arm cuff correctly, then take another
measurement. Refer to section 4 of instruction manual
Remain still and sit correctly during a measurement.
If the "
" symbol continues to appear, we recommend
you to consult with your physician.
EN
.
.
.
EN4

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Hem-7121j-e

Tabla de contenido