ERROR INDICATORS AND TROUBLESHOOTING TIPS
ERROR INDICATORS
SYMBOLS
After several seconds
After several seconds
After several seconds
CAUSE
Wrist movement
during the
measurement.
Arm or body
movement during the
measurement.
Talking during the
measurement.
Excessive arm or
body movement
during measurement.
69
CORRECTION
Turn the unit off.
Refer to "Taking a
Measurement". Take
another measurement
and do not move until
the measurement
results display.
Turn the unit off. Refer
to "Taking a
Measurement". Take
another measurement.
Do not move or talk until
the measurement results
display.
Turn the unit off.
Refer to "Taking a
Measurement". Take
another measurement
and do not move until
the measurement
results display.