5.2.3 Changing the Range Limits
The given limits for the input range can be en-
tered as is, or adjusted.
Parameter
Possible range of values
0 .. 10 V
–0.500 ... 10.500
2 .. 10 V
01.500 ... 10.500
-10 .. +10 V –10.500 ... 10.500
0 .. 20 mA
–1.000 ... 21.000
4 .. 20 mA
03.000 ... 21.000
If the measured signal falls below or exceeds the
programmed value, then the display alternates
between
and the measured value or
between
and the the measured value.
Lower limit
Menu
<–> Selection
press the
key to accept the selection
Upper limit
Menu
<–> Selection
press the
key to accept the selection
www.kuebler.com
Parameter
Possible range of values
–0.500 ... 10.500
01.500 ... 10.500
–10.500 ... 10.500
–1.000 ... 21.000
03.000 ... 21.000
Example: –5.000
Select digit
Set digit
Select digit
Set digit
Example: 9,000
Select digit
Set digit
Setting values out of the range is not possible. It
is only possible to continue with the set-up,
using the
key, when the settings are
correct.
When the signal drops below the value set here,
then the signal alternates with the message
Under-range: if the signal is less then –13,60 V
than
appears in the display.
Current values < 0.0 mA will not be
measured.
When the signal exceeds the value set here,
then the signal alternates with the message
.
Over-range if the signal is higher than 11.00 V or
21.5 mA, then
appears in the display.
11