Descargar Imprimir esta página

Health O Meter PRO PLUS 1100KL Manual Del Usuario página 33

Balanzas electrónicas
Ocultar thumbs Ver también para PRO PLUS 1100KL:

Publicidad

Idiomas disponibles

Idiomas disponibles

Name
ESC
character
Response
P
Diagnostics(1)
A
Diagnostics(2)
Z
Control (set a
C
value)
Get subject
G
record data and
readings
Set subject record
S
data
Clear readings
X
Clear subject
Y
record data
Name of Control
Auto Hold Time
Auto Off Time
Sleep Time
Keypad Volume
Stable weight Volume
Alarm Volume
Display of Date/Time
Backlight
Contrast
Weight
Display
decimals(d) or Pounds and ounces (p)
Tare Weight
Unit Of Measure (Metric or constitutional)
Positions left in RAM
Table A - Scale Global Values List and Identifiers
Name of component to test
Load Cell A
Load Cell B
Load Cell C
Load Cell D
Battery
PC Communication (USB)
Write to RAM
Read from RAM
Table B - Components to test
of 10 seconds like this: <ESC>PAHT=10<ESC>E )
ESC value with
parameters
Pccc=ccccccc This will be the scales response to the request mentioned above
Accc
This is the request for a diagnostic test on certain parts of the
scale (such as battery life, loadcells, keypad, LCD etc.). (See
Table B)
Zccc
This will be the response of the diagnostics done on the scale.
Values will include error codes to indicate what is wrong with the
scale, or all zeros to indicate that all is well.
Cccc=cccccc
This is to set the value of the scale's global settings such as
backlight, auto hold time, keypad beep volume etc. (See Table
A) (i.e. <ESC>CAHT=20<ESC>E will set the Auto Hold Time to
20 seconds)
G
This will return all the information from both the subject record
data as well as the readings stored in the readings table.
S
This will set the subject record data for a specific patient.
Immediately following this will be the value that is set. (i.e.
<ESC>S<ESC>I1234<ESC>T200<ESC>Um<ESC>E will set
the TARE for patient ID1234 to 200kg)
X
This will clear all the values stored in the readings table on the
scale. Only the subject record data will not be deleted.
Y
This will clear the subject record data for a specific patient. The
patient ID will follow the Y immediately (<ESC>Iccccccccccc). If
no ID is given, ALL the data will be cleared.
Mode
(fractions(f),
Identifie
r
LCA
LCB
LCC
LCD
BAT
PCC
WRM
RRM
Description
Identifier
Unit
AHT
nn (Seconds)
AOT
nn (Minutes)
STM
nn (Seconds)
VKE
nn (01..10)
VSW
nn (01..10)
VAL
nn (01..10)
DAT
nn (01 or 10)
BAC
nn (01..10)
CON
nn (01..10)
WDM
c (f,d,p)
TAR
nnn.nn (Kg)
UOM
c (m or c)
RAM
nnnn (0..max)
Error Code
E3A
E3B
E3C
E3D
E4L(Bat Low, but still usable) or
E4U (Bat Low and Unstable)
E05
E08
E09
31

Publicidad

Solución de problemas

loading

Este manual también es adecuado para:

Pro plus 2650klPro plus 2700kl