The different tables on the scale:
There will be two data tables on the scale. The first will store the subject record data (patient-related
data that does not change) and the second will store the data recorded during the readings made.
Subject record data
Fields: ID, TARE, HEIGHT
Readings
Fields: Index (referring to the ID from the Subject Record Table), Weight, Date-time
Sample and explanation of esc protocol
1. When the scale measures the weight of a patient and sends this over the
communications line to the PC, it will look like this:
<SCALE>
<ESC>R<ESC>I1234567890<ESC>W90.98<ESC>H178.7<ESC>T5.7<ESC>B28<ESC>Nm<ESCE
2. When the PC requests the values of the stored weighings, it will look like this:
First the PC will ask the scale to send the data. We do this with the "Get Subject Record Data and
Readings" command.
<SCALE>
<ESC>G<ESC>E
Then the scale will reply with the stored data.
Direction of communication
Direction of communication
Get subject data
EOP
32
<PC>
Reading (Command)
Patient ID = 1234567890
Weight = 90.98
Height = 178.70
TARE = 5.7
BMI = 28
Unit of measure = Metric
EOP