Performing A Control Solution Test - Fora Pro Voice V8 Manual Del Propietário

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Performing a Control Solution Test

WARNING:
When doing the control solution test, you have to mark it so that the test result
will NOT mix with the blood glucose TEST RESULTS stored in the memory. Failure
to do so will mix up the blood glucose test results with the control solution test
results in memory.
18
1. Insert the test strip to turn on the
meter
Insert the test strip into the meter. Wait
for the meter to display "
Voice Guide
Thank you for using FORA product. Please
relax during measurement. Please apply
blood onto the test strip.
2. Press Main button to mark this test
as a control solution test
With "
" displayed, the meter will not
store your test result in memory. If you
press the Main button again, the "
will disappear and this test is no longer a
control solution test.
".
"

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido