Important Control Solution Information
● Use only our control solutions with your meter.
● Do not use the control solution beyond the expiration date or 3 months
after first opening. Write the opening date on the control solution vial
and discard the remaining solution after 3 months.
● It is recommended that the control solution test be done at room
temperature 68°F to 77°F (20°C to 25°C). Make sure your control
solution, meter, and test strips are at this specified temperature range
before testing.
● Shake the vial before use, discard the first drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
● Store the control solution tightly closed at temperatures between
35.6°F to 86°F (2°C to 30°C). Do NOT freeze.
35