3. Apply control solution
Shake the control solution vial thoroughly before use. Squeeze out a
drop and wipe it off, then squeeze out another drop and place it on the
tip of the vial cap. Hold the meter to move the absorbent hole of the test
strip to touch the drop. Once the confirmation window fills completely,
the meter will begin counting down. To avoid contaminating the control
solution, do not directly apply control solution onto a strip.
Out-of-range results
If you continue to have test results fall outside the range printed on the
test strip vial, the meter and strips may not be working properly. Do NOT
test your blood. Call customer service for help.
NOTE:
• The control solution range printed on the test strip vial is for control
solution use only. It is not a recommended range for your blood glucose level.
• See the Meter Maintenance section for important information about your
control solutions.
4. Read and compare the result
After counting down to 0, the control
solution test result will appear on the
display. Compare this result with the
range printed on the test strip vial and it
should fall within this range. If not, please
read the instructions again and repeat
the control solution test.
100~165
(100−165 mg/dL = 5.6−9.2 mmol/L)
The blood glucose is (number) milligram
per deciliter.
Voice Guide
19