8.4.2 Error messages
Error message
[Lo]
[uLoad]
[oLoad]
[no 0.00]
„-----"
„-----"
8.5
Maintenance
To ensure correct measurement, maintenance and repair should only be carried out
by authorised personnel.
To prevent the intended level of accuracy the product must be set up carefully and services
regularly. We recommend having it serviced every 3 to 5 years depending on how often the scales
are used.
8.6
Recalibration
According to the national regulations of the legislature, only authorized companies or authorized
personnel can perform recalibration. On the CE mark, you can find the year of initial calibration
next to the notified body (0122).
Recalibration must be carried out when:
-
the calibration counter reading displayed when the scale is switched on does not match the
number marked on the valid calibration counter mark,
-
one or more security marks were damaged,
-
after repair of a calibrated scale,
-
after expiry of the period, set by the national regulations for recalibration.
8.7
Storage and Transport Conditions
Keep all original packaging materials and components for eventual return of the scale to avoid
damage during transport; these are not covered by the warranty.
To avoid damage disconnect all cables before transport.
Seite 32 von 112
Description
Empty battery.
Underload (-20d)
Overload (-9d)
No zero point available.
Negative weight values (e.g. in
the tare function) are not
displayed.
Scale is loaded too slowly
during the HOLD function.
Troubleshooting
Insert new batteries.
Use the scale in mains
operation.
Switch the scale off and on
again.
Unload the scale. Zero range or
weighing capacity exceeded
Zero the scale.
Load the scale.
Unload the scale. Wait until
>0< is shown in the display.
Load the scale again.
M118000+M118000-01-210305-Rev002-UM-en