Installation

Inspect the controller for any shipping damage. The controller is
designed to be installed into a rack. All the connections are fitted to
J2, J3, J4 and J5 connectors. Should the controller be connected to a
host computer via the RS-232 interface, a suitable cable must be
prepared. In the following paragraphs are detailed the input/output
signals.
The PCB installed into the customer system must be positioned so that cold air
NOTE
(forced or natural convection) can flow through the PCB components.
J3 Connector
Figure 7
The signals of J3 connector are the following:
TwisTorr 74 FS PCB 24 Vdc Controller User Manual / 87-901-056-01
J3 – Power Connector
Pins 1/2 Power supply 24. Vdc (20 to 32 Vdc) (pin 2 positive).
Pins 4/3 Kl relay contacts. The relay switches when the output
frequency overrides the set threshold (1117 Hz). The threshold
overriding is indicated also on pin 6 of J4 connector.
Technical Information

Installation

179/209