6. Faults
Dial status
40
60
10
15
Power
20
5
316L
316L
20
4
mA
%
0
40
60
10
15
Power
20
5
316L
316L
20
4
mA
%
0
40
60
10
15
Power
20
5
316L
316L
20
4
mA
%
0
1) In accordance with recommendation NE43 of NAMUR (international user association of automation technology in process
industries)
WIKA operating instructions indicator for panel mounting, model 910.70
Description
Power LED flashes red.
Pointer is above the full scale
value.
Analogue output: > 20 mA
80
100
Power LED flashes red.
Pointer is below the scale.
Analogue output: < 4 mA
80
100
Power LED does not light up.
Pointer is below the scale.
Analogue output: > 21 mA
80
100
Operating status
The sensor signal is between
20.5 ... 21 mA
.
1)
Error signal is suppressed for a
few seconds.
Check sensor.
Sensor signal is < 3.6 mA
Error signal is suppressed for a
few seconds.
Check sensor cable and sensor.
Check power supply
EN
1)
13