7.1.2.2. Use of combination valve pressure outlet
Ensure that the accessory IS NOT connected to the pressure outlet.
Open slowly the combination shut-off valve by turning slowly the combination control knob
"shut-off valve & fl ow selector" in anticlockwise direction to ON position.
Connect the accessory to the pressure outlet.
Before connecting any accessory to the pressure outlet make sure that the patient is not con-
nected and the accessory outlet is secured.
If pressure outlet is to be connected to a medical device that requires high gas fl ow (for
instance pulmonary ventilator that requires gas fl ow 100 l/min at the minimum pressure 2.8
bar), compare the required fl ow of connecting medical device with pressure and fl ow charac-
teristics of the combination valve stated in appendix Nr. 1. To assure su cient performance
(pressure and fl ow characteristics of the combination valve ) the medical device should not be
used if the content indicator of Manómetro Digital Display enters the red zone..
Sudden opening could result in a danger of fi re or explosion arising from oxygen pressure
shocks. Insu cient opening of the shut-o valve could reduce actual fl ow delivered.
If a pressure outlet as well as a fl ow outlet are part of the combination valve do not use them
simultaneously, especially if pressure in the cylinder is below 50 bar, in could adversely a ect
the outlet parameters of the combination valve .

7.1.3. AFTER USE

1. Turn the combination Combination Control Knob 'Shut-off valve & Flow Selector'
clockwise direction to the OFF position. Do not use excessive force. Do not use excessive force
(Maximum recommended closing torque is 3 Nm).
2. Vent pressure from the connected devices.
3. Disconnect all connected devices from user outlets.
4. Refi t pressure outlet and fl ow outlet protection cap (if available).
(Before refi tting the caps, ensure they are clean).
8. ACCESSORIES
Accessories connectable to fl ow outlet:
hose connected with mask, cannula or humidifi er.
Accessories connectable to pressure outlet:
low pressure hose (working pressure >10 bar), fl owmeters, Venturi suction ejectors, lung
ventilators.
Other user accessories:
bed hanger, humidifi er holder.
Accessories for fi lling stations:
fi lling adaptor.
Before connecting any accessory or medical device to the combination valve, always check
that they are fully compatible with connection features & performances of the product.