6.4. DIGITAL DISPLAY OVERVIEW
K. Warning indicator
J. Bluetooth active
I. Flow rate (L/MIN)
EN
H. Gas content indicator
A.
Gas content indicator (H)
The segmented bar indicates how full the cylinder is according to the scale.
The content scale is divided into 13 segments as follows:
Between 0 (Empty) and ¼ the scale is divided into four segments –
Between ¼ and ½ the scale is divided into three segments –
Between ½ and 1 (Full), the scale is divided into six segments –
B.
Cylinder content and time remaining indication (N1 and N2)
This gives the remaining time available for treatment at the current fl ow rate.
This is only displayed when the valve is open and gas is fl owing:
When the fl ow rate is altered, the actual gas fl ow will change immediately, but the treatment
time remaining will need to be recalculated. During the remaining-time calculation the cylinder
content in litres (LTR) 'N1' is visible and the 'L/MIN' icon, label 'I' on digital display, blinks to
indicate that the remaining time calculation is in progress
If the display shows "--:--" (without the Alert icon 'L' being displayed) then it indicates a system
fault. Please contact the gas package owner.
C.
Cylinder gas content in litres (N1)
If cylinder is not in use or the remaining treatment time calculation
is not underway the cylinder content is indicated in standard gas litres.
When a cylinder is in storage or there is no cylinder movement, the digital display will move
into sleep mode. The system will automatically re-awaken if a fl ow is detected or the cylinder
is gently moved.
D.
Remaining treatment time at set fl ow rate (N2)
If cylinder is in use and the remaining-time calculation is complete,
the remaining treatment time will be indicated in "hours : minutes".
E.
Bluetooth Active (manufacturer use only) (J)
When displayed, this icon indicates that the Bluetooth connection is active.
This is for use by the manufacturer during service operation.
1
/
/
L
M
K
I
J
N2
8/56
L. Alerts
M. Battery icon
N1. If LTR visible,
number displayed
below is contents
in standard litres
N1
N2. If HR:MIN
visible, number
displayed below
is remaining treat-
ment time,
at set fl ow rate
Red area
Yellow area
Green area