6. Click OK.
7. Select the database from the USB flash drive. Go to USB > Database > Backup 20XX.XXX >
TEIS_DATA.FDB
8. Close the software and restart the PC.
Section 5 Startup
Items to collect: flow meter
The first time the analyzer is used, do the steps that follow:
1. Set the PC to on.
2. Set the power switch of the analyzer to on.
3. Open the gas supply. Make sure that the gas pressure is less than 8 bar (116 PSI).
4. Start the TEIS2 software. Login with service engineer credentials.
5. Wait until all of the devices in the software show as connected.
6. Use the flow meter to examine the gas connections. Then, install the correct item.
a. Connect a flow meter to the lower right connection of the humidifier.
The flow must be 150 mL/min ± 5 %.
b. Fill the humidifier with ultra-pure water. Carefully connect the humidifier.
Note: Make sure to push the humidifier in a straight line into the connector or damage to the connector can
occur.
c. Connect the flow meter to the carrier gas tubing in the top of the furnace.
The flow must be 150 mL/min ± 5 %.
d. Connect the carrier gas tubing at the top of the furnace. Install the top cover of the furnace.
e. Connect the flow meter to the middle connection of the IC vessel.
English 15