DCM4010FLEX
Backlight
Data Hold
Replacing Batteries
Caution
Remove Meter from conductor and turn off
power before opening the battery cover.
Environmental Specifications
Indoor Use.
Pollution Degree : 2
Maximum operating altitude : 2000m (6562 ft)
Operating temperature & Relative Humidity :
0°C - 30°C, ≦80%RH
30°C - 40°C, ≦75%RH
40°C - 50°C, ≦45%RH
Storage temperature :
-20°C - 60°C, 0 to 80% RH (no batteries).
Temperature coefficient :
0.2 x (Specified accuracy) / °C, < 18°C, > 28°C
Shock vibration : Sinusoidal vibration
MIL-PRF-28800F (5-55 Hz, 3g max.)
Drop Protection : 4 feet drop to hardwood on
concrete Floor
Electrical Specifications
Accuracy is given as ±(% of reading + counts
of least significant digit) at 23°C ± 5°C, with
relative humidity Less than 80% R.H.
Accuracy is specified for a period of one year
after calibration.
Position Error of Clamp :
Accuracy and position error assumes
centralized
primary conductor at optimum position, no
external electrical or magnetic field, and within
operating temperature range.
sFlex-10
15
Distance from
25
Optimum (mm)
35
AC Current
OL
Range
Resolution
Reading
30.00A
33.00A
300.0A
330.0A
3000A
3300A
LCD displays 0 counts when the reading < 10
counts.
Frequency Response : 45 - 500Hz (Sine Wave)
AC Function
ACA specifications are ac coupled, true R.M.S.
For non-sinusoidal waveforms, Additional
Accuracy by Crest Factor (C.F.):
Add 3.0% for C.F. 1.0 - 2.0.
Add 5.0% for C.F. 2.0 - 2.5.
Add 7.0% for C.F. 2.5 - 3.0.
03
EN
Error
±2.0%
±2.5%
±3.0%
Accuracy
(45Hz~500Hz)
0.01A
±(3.0% + 5D)
0.1A
±(3.0% + 5D)
1A
±(3.0% + 5D)