inbody 270 Manual De Instrucciones página 28

Ocultar thumbs Ver también para 270:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

3. Rear Panel
Speaker: Provides audible indication for test in progress, test complete and successfully saved setting changes.
9-pin COM serial port (Female, RS-232C): Used for connecting the InBody270 to LookinBody installed
on a computer, a stadiometer, a blood pressure monitor and SD400.
The InBody270 can be connected to LookinBody installed on a computer using one of the ports
*
Only compatible with a InBody Stadiometer, Blood Pressure Monitor, and SD400.
*
LAN port (10T Base): Used for connecting the InBody270 to LookinBody installed on a computer.
* The InBody270 can be connected to LookinBody installed on a computer using one of the ports
USB SLAVE port: Used for connecting the InBody270 to LookinBody installed on a computer.
* The InBody270 can be connected to LookinBody installed on a computer using one of the ports
USB HOST port: Used for connecting to a printer, a barcode reader, or a USB Thumb Drive.
Power Inlet: Used for connecting the power adapter.
* Always use the adapter supplied by InBody Co., Ltd..
Power switch: Used to turn the device on/off.
UTION
NOTE
참 고
주 의
Do not touch signal input, signal output or other connectors, and the patient simultaneously.
External equipment intended for connection to signal input, signal output or other connectors, shall comply
with relevant IEC Standard(e.g., IEC60950 for IT equipment and IEC60601-1 series for medical electrical
equipment). In addition, all such combination-system-shall comply with the standard IEC60601-1 and/or
IEC60601-1-1 harmonized national standard or the combination. If in doubt, contact qualified technician
or your local representative.
28
DANGER
Warning
참 조
주 의
, or
.
❷,
, or
.
❷,
, or
.
❷,

Publicidad

Tabla de contenido
loading

Tabla de contenido