Charging Function For Gas Detectors Of The X-Am Series (Optional) - Dräger X-dock 5300 Instrucciones De Uso

Ocultar thumbs Ver también para X-dock 5300:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 51
3. Screw pressure regulator valve on full test gas cylinder with
identical test gas concentration.
4. Slowly open the test gas cylinder valve.
To replace a test gas cylinder with a test gas cylinder with other
test gas concentration:
1. Close empty test gas cylinder valve.
2. Unscrew the pressure regulator valve from the test gas
cylinder.
3. Screw pressure regulator valve on full test gas cylinder with
changed test gas concentration.
4. Slowly open the test gas cylinder valve.
5. Reconfigure the corresponding test gas inlet, see
Chapter 4.6 on Page 24.
6.3
Charging function for gas detectors of
the X-am series (optional)
Gas detector
Charging module
X-am 2500
X-am 125+ module
X-am 5000
X-am 5600
X-am 3500
X-am 8000+ module
X-am 8000
1
With the rechargeable battery completely empty.
2
A new NiMH supply unit reaches its full capacity after 3 full
charge/discharge cycles.
The station offers 2 different charging functions:
Charge after 15 minutes of inactivity
Direct charging of a switched off gas detector
To charge a gas detector using a X-am 125+ module after a
test:
1. Place the gas detector in the X-am 125+ module.
2. Close the module flap.
The instrument will be detected automatically.
After the instrument is inserted, the charging status is
displayed for approx. 5 seconds via the charging status
LED.
The charging function starts automatically approx.
15 minutes after the last test.
The station additionally offers an option for directly charging
switched-off gas detectors with no waiting period. If this option
is activated, the gas detectors will not be automatically
switched on when inserted into the modules. The X-am 125+/
8000+ modules then start charging directly.
To charge a switched-off gas detector directly using an
X-am 125+/8000+ module:
1. Activate the option Do not switch on device on the station
(see Technical Manual).
2. Insert the switched-off gas detector in the X-am 125+/
8000+ module.
3. Close the module flap.
The gas detector will be detected automatically and
charged directly.
Dräger X-dock 5300/6300/6600
If an error occurs:
!
i
i
Overview of charge status LEDs
1
Charging time
Colour
2
approx. 4 h
green
green
red
approx. 9-10 h
6.4
The maintenance station can be operated in the following
connection modes:
6.4.1
1. Select
2. Select XDM.
3. Establish a secure network connection (see technical
6.4.2
Prerequisites:
1. Select
2. Select GDC.
3. Confirm the default network settings or change them if
4. Confirm information.
5. Enter tenant PIN and confirm.
Remove from the module and re-insert.
If this does not correct the error, have the module repaired.
CAUTION
With X-am 125+ charging modules only: A short-circuit
in the charging contacts in the modules, e. g. due to
metal objects falling into the device, will not damage
the station but should be avoided due to a potential risk
of overheating and error displays on the module.
NOTICE
With X-am 8000+ charging modules only: Metal
objects in the charging cradle can have a negative
impact on the charging function, and may lead to an
error or prevent that the gas detector is charged.
Status
on continuously
flashing
flashing
Changing the connection mode
Single mode
X-dock Manager mode (XDM)
Gas Detection Connect mode (GDC)
Connecting the maintenance station to XDM
> System configuration > Change connection
mode.
The maintenance station will be shut down and restarted in
XDM mode.
manual).
Connecting the maintenance station to GDC
GDC is available.
Tenant PIN is available. The PIN can be requested via the
Dräger website.
> System configuration > Change connection
mode.
necessary (see technical manual).
Information about commissioning is displayed.
After a successful check, the maintenance station will be
shut down and restarted in GDC mode.
Maintenance
Meaning
Charge status 100 %
Battery charging.
Charging error
29