3. If required, configure other gas inlets in the same way.
When using a gas cylinder from another manufacturer:
1. Create or delete test gas component.
Create a new test gas component with
Delete the current test gas component with
NOTICE
i
i
Deleting all test
information of the test gas inlet.
2. Select gas.
3. Enter the gas concentration.
4. Select gas units.
5. If required, create additional test gas components.
6. The following information can optionally be given:
Part no. of the gas cylinder
Batch number of test gas cylinder
Expiry date of the gas cylinder
7. If required, enter Further options.
8. For more information see X-dock 5300/6300/6600
Technical Manual.
Dräger X-dock 5300/6300/6600
.
.
gas components deletes all
4.6.1
Settings
> Station gas configuration > Settings.
1. Select
Fresh air inlet setting
Expert settings
Gas cylinder level indicator
To set the fresh air input:
1. Select Fresh air.
2. Select desired setting:
Supplied by pump - fresh air inlet (default setting)
Supplied by compressed air input - compressed air
inlet
3. Confirm selection with OK.
The following settings can be made in the expert menu:
Ignore max. concentration for quick bump test
Set test behaviour for missing test gases
To ignore the concentration limit specified by Dräger for the
quick bump test:
1. Select Ignore max. conc. for BTQ.
2. Activate checkbox (default setting: deactivated).
3. Confirm selection with OK.
Activating this feature allows users to use higher test gas
concentration for the quick bump test than recommended by
Dräger.
WARNING
!
This feature must only be activated by trained and
qualified personnel since wrongly selected test gas
concentrations may lead to a positive test result
although the gas detector's warning comes too late.
To set the test behaviour with missing test gases:
1. Select Cancel the test on missing gas.
2. Activate checkbox (default setting: activated).
3. Confirm selection with OK.
This function can be used to set whether or not a test or
calibration is carried out when a required test gas is not
connected.
WARNING
!
If this function is deactivated, the corresponding
channel is not tested or calibrated.