NOTICE
i
i
Single mode is activated by default.
Several tests can be started and executed in parallel in
the Single mode.
A failure of a LED, horn or vibration test results in a
negative evaluation of the overall test, and thus in the
locking of the respective gas measurement system.
A testing of the sensor reserve is only performed with
sensors that support this function. The results are
displayed under test details and give information about
the state of the sensor.
NOTICE
i
i
If an X-am 8000 remains off for longer than 21 days
and is not charged, deep sleep mode is activated
automatically. In deep sleep mode, the gas detector
can no longer be switched on using the Dräger
CC-Vision PC software or the Dräger X-dock. In this
case, the gas detector must be switched on manually.
The following tests are preconfigured (for more information see
the X-dock 5300/6300/6600 Technical Manual.):
Quick bump test including checking the alarm
Test 1: QT
elements.
Extended bump test incl. zero-point check and
Test 2: EXT
checking the alarm elements.
Test 3: CAL
Calibration and checking the alarm elements.
1. If necessary, open the test gas cylinders.
2. If necessary, switch on the X-dock.
3. Conduct a visual inspection of the instruments (see
Chapter 5.1 on Page 26).
4. Insert the instruments into the modules (see Chapter 5.2
on Page 26).
If Single mode is activated:
The default test is started automatically when the module
flap closes.
The Status LED flashes blue.
The individual test phases are displayed.
If Favorites mode is activated:
1. Select the desired test from favourites bar.
2. If the test requires a higher authorisation level, log in at the
station with the respective authorisation level (see
Chapter 4.5 on Page 24).
Test will be started automatically.
The status LED flashes blue.
The individual test phases are displayed.
If Test schedulermode is activated:
If necessary, log user out on the station (see Chapter 4.5
on Page 24).
The preset test is performed according to the configured
time schedule.
Dräger X-dock 5300/6300/6600
Test passed:
Confirmation is shown on the display.
The Status LED flashes green.
If required, select the desired instrument field for additional
information (e.g. test date and performed tests).
Remove the instrument from the module.
WARNING
!
Prior to each use of the gas detector, check whether it
is switched on and in measuring mode after removing
it from the X-dock. Otherwise the user may carry a
device that is switched off!
NOTICE
i
i
After the favourite has passed through, cross-
sensitivities may cause some sensor combinations to
drift around the zero point until shortly after the gas
detector is removed from the station. This effect can be
prevented with the Purge option (see Technical
Manual).
Test passed, with reservations:
Use
0 0
01033286.eps
0 0
01133286.eps
27