The application tiles can be configured as follows:
■
Enlarge/shrink tile (touch gesture)
■
Move tile (touch gesture)
■
Rename tile (entry via virtual keyboard)
Note
The edit mode must be deactivated and the user mode activated to then
start applications, see Chapter 7.5.
7.4
Device Status/Device Connection
Fig. 1
In this area, status information on the OrthoPilot
nected periphery devices OrthoPilot
wireless footswitch are displayed.
Note
®
If the OrthoPilot
4 wireless footswitch is note being used, the status dis-
play can be deactivated upon first commissioning.
►
Display on information on device and options for function tests and
error handling: select the button of the device.
There are two types of messages:
Symbol
Designation
Warning
Malfunction
®
7.4.1
OrthoPilot
4 panel PC
Information, such as article number, serial number and battery status of
®
the OrthoPilot
4 panel PC is displayed via the "Panel PC" button.
Battery status
If the panel PC is being powered during operation via the integrated USV,
a warning is displayed. The battery of the USV is designed so that the panel
PC can be provided with power for at least 2 minutes in the event of a
power outage. If there is a power outage during application workflow, this
will be displayed in the application (for additional information see the
instructions for use of the applications).
If the cause of the voltage drop cannot be detected and rectified immedi-
ately, the system will shut down automatically when the battery runs low.
Message
Message displayed
Warning: devices
Device can be used under certain framework condi-
is running on
tions.
emergency
Read the warning and follow the instructions.
power
Error: critical
Device cannot be used.
battery status
Alleviate the error with the troubleshooting dialog.
►
Check that the mains plug has not been pulled out inadvertently.
® Elite
When the OrthoPilot
Hardware System is started up, the USV is auto-
matically function tested. If a bridging of voltage interruptions cannot be
guaranteed (such as due to a dead of defective USV battery), a warning is
displayed.
►
Only use the device when the battery has bee charge to a point that the
USV functionality can be guaranteed during operation.
®
4 panel PC and the con-
®
4 wireless camera and OrthoPilot
Description
Device can be used under certain frame-
work conditions.
Read the warning and follow the instruc-
tions.
Device cannot be used.
Alleviate the error with the trouble-
shooting dialog.
®
7.4.2
OrthoPilot
®
4
Fig. 2
Information, such as article number, serial number and firmware status is
displayed via the "Footswitch" button. A function test can also be per-
formed.
Commissioning (only necessary when connecting a new foot switch)
►
Switch on the OrthoPilot
the bottom to the "ON" position "I".
Note
The on/off switch is provided solely as a transport safeguard and can remain
in the ON "I" position after initial startup.
►
Select the "Footswitch" button.
►
Press the right and left pedal at the same time to activate the connec-
tion mode.
►
Select the search button in the dialog window.
A list of the OrthoPilot
Note
If multiple devices are displayed, the foot control can be identified by the
serial number. The serial number is located on the type plate on the bottom
of the foot control.
►
Select the OrthoPilot
ber.
A function test can be performed once connected.
►
Select the "Keyboard" dialog window and press the pedals in succes-
sion.
The activation of the pedals will be displayed accordingly.
®
The OrthoPilot
4 wireless footswitch is now ready to use and can be
placed under the operating table in the user's work area.
Note
The range of the radio connection between the OrthoPilot
switch and theOrthoPilot
4 wireless footswitch (optional)
®
4 wireless foot switch: turn the ON/OFF on
®
4 footswitches found appears.
®
4 wireless footswitch with the right serial num-
®
4 panel PC unit is max. 5 m.
en
®
4 wireless foot-
9