Reader Unit Set
MALFUNCTIONS DISPLAYED WITH AN ERROR MESSAGE
Message displayed
Battery flat - Auto off
Battery voltage incorrect
– use original power supply unit
Low battery voltage
Flashing of crossed-out antenna
symbol
<Info> button: Antenna faulty
Flashing of crossed-out antenna
symbol
<Info> button: Antenna not
plugged in
Flashing of crossed-out antenna
symbol
<Info> button: No communication
Flashing of crossed-out antenna
symbol
<Info> button: SD card has been
removed. Measurement possible
Dataset defective!
<Info> button: File cannot be
opened
Continuous key activation Keypad
error
Pressure readings out of range
Problem with input voltage
Wrong SD card inserted! Remove
card!
<Info> button: Measurement with-
out data storage possible - or -
Insert SD card with correct ID
XXXXXXXXXX!
Cause
Battery capacity exhausted (0%)
Battery voltage of the Reader Unit
is too low
Battery voltage too low
antenna faulty
Antenna not connected at the
start of measurement - or -
Antenna was disconnected during
measurement
Data recording interrupted during
continuous measurement (inter-
ruption of telemetric coupling)
SD card was removed during con-
tinuous measurement
Validation of file failed
Persistent pressing of a button >
60 seconds
Pressure data measured from
implant not plausible - physiologi-
cally meaningless data
The voltage of the plug-in power
unit is too high
Measurement starts without SD
card. During measurement, an SD
card not designated for use with
that implant is used during mea-
surement
INSTRUCTIONS FOR USE | GB
Error detection / troubleshooting
All data are saved after 2 min.
Reader Unit switches off automati-
cally. Original plug-in power unit.
Reader Unit switches off automati-
cally after 20 seconds. Connect the
original plug-in power unit.
Backlight switches off after 3 sec-
onds. Connect the original plug-in
power unit. No interruption to ongo-
ing measurements.
Deactivate and reactivate device.
If the error persists, contact Tech-
nical Service.
Connect antenna: measure-
ment starts again - or - Connect
antenna: measurement continues.
Measurement resumes auto-
matically once communication is
restored.
Insert SD card. Start measurement
again.
File cannot be opened. Try again.
Release the key.
Measurement continues. If the
error persists, contact Technical
Service.
Reader Unit switches off automati-
cally after 20 seconds. Use original
plug-in power unit.
Insert the matching M.scio SD card
into the Reader Unit. The implant
(see patient data card) is associ-
ated with the respective SD card
via the identification number (ID).
45