Procedure to be performed:
1) Turn the locking ring (1) fully anti-clockwise and remove
the head.
2
1
2) Fit the adapter for Ivory heads and contra-angles to the
oil spray can. Fit the head (A) in the adapter (B). Hold the
head in place. Spray for about 5 seconds.
A
B
42
Eng
3) Apply the adapter to the back of the handpiece. Hold
the handpiece in place and spray for about 5 seconds.
4) Replace the head in the handpiece.
CAUTION!
Incorrect or inadequate spindle cleaning can cause the
spindle to loosen and accidentally release the bur during
functioning.
CAUTION!
After lubricating, hold the handpiece with the chuck
facing down. Operate the handpiece for 30 seconds to
expel any excess oil. If you notice any dirt coming out,
repeat the entire cleaning and lubrication process.