Calibrate the gas detector

5.6.1 Notes on calibration

WARNING
Incorrect measured values
Incorrect calibration means that alarms may not be triggered or only with a
delay.
► Always carry out the clean air/zero calibration prior to the span calibration.
If the measured or calibration gas is changed, the affected channel must be
calibrated.

5.6.2 Performing a fresh air calibration

Observe the following instructions for fresh air calibration:
● To improve accuracy, a fresh air calibration must be carried out if a zero
deviation exists.
● For the fresh air calibration, the display on the DrägerSensor XXS O
XXS O
PR is set to 20.9 Vol%.
2
Prerequisites
The fresh air calibration can only be carried out if at least one sensor
supports fresh air calibration.
The fresh air must be free of measured or interfering gases.
The gas detector is switched on and warm-up phases 1 and 2 are complete.
To carry out a fresh air calibration:
1. Switch on the gas detector.
2. Call up fresh air calibration (depending on configuration):
Via the Quick menu:
● Press
● Select and confirm fresh air calibration
Via the menu:
a. Press
b. Enter and confirm the password.
c. Select and confirm fresh air calibration
 All measurement channels involved in the fresh air calibration flash.
®
X-am
2800
3x in measuring mode.
for approx. 4 s in measuring mode.
Maintenance
.
.
|
en
and
2
55