Device Settings; Factory Settings - Dräger X-am 2800 Instrucciones De Uso

Ocultar thumbs Ver también para X-am 2800:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 101
en
|

Device settings

6
Device settings
Only trained and qualified personnel may change the device settings.
For more details, refer to the technical manual.
6.1

Factory settings

Different settings can be selected to meet customer requirements when
ordering. The setting can be checked and changed with the Dräger CC-Vision
PC software.
The changed parameter settings must be checked after being transferred to
the gas detector to ensure that the values have been transferred correctly.
Parameters that cannot be viewed on the gas detector must be read out and
checked after being changed using the Dräger CC-Vision PC software.
Feature
Fresh air calibration without password On
Bump test without password
Operation signal (acoustic)
®
Bluetooth
Switch off allowed
1)
Capture range
2)
LEL factor
3)
CH
(methane)
4
H
(hydrogen)
2
C
H
(propane)
3
8
STEL
TWA
Configuration type alarm thresholds
Alarm A1
60
Setting
On
Off
Off
On
On
4.4 Vol% (corresponds to 100 %LEL)
4.0 Vol% (corresponds to 100 %LEL)
1.7 Vol% (corresponds to 100 %LEL)
STEL feature – inactive; average dura-
tion = 15 minutes
TWA feature – inactive; average dura-
tion = 8 hours
ATEX compliant
Acknowledgeable; non-latching, pre-
alarm, increasing measured value (for
O
sensor, decreasing measured value
2
as well)
®
X-am
2800

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Mqg 0200

Tabla de contenido