ProMinent CTE 1-mA Instrucciones De Servicio página 33

Sensor de cloro total
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Assembling
the sensor
Insertion
of sensor
into the in-line
probe housing
Note also the instructions and safety guidelines in the operating instructions
of the in-line probe housing.
DLG III
DGM
®
ProMinent
Place the electrode shaft vertically onto the filled membrane cap.
Do not cover the vent hole below the rubber seal with your fingers.
Screw on the membrane cap by hand as far as possible. There should
not be a visible gap between the electrode shaft and membrane cap.
When you screw on the cap, excess electrolyte will leak out through the
hole below the rubber seal. (see fig. 3).
Wipe away any electrolyte with a soft paper tissue or similar.
Rinse the nozzle thoroughly with a clean, warm water jet so that no
electrolyte remains on the inside and outside of the nozzle.
IMPORTANT
• Depressurise the system before inserting the sensor into
the in-line probe housing. Close the stop valves before
and after the in-line probe housing.
• Insertion and withdrawal of the sensor into or out of the
in-line probe housing should be carried out slowly.
• Do not exceed the maximum operating pressure of 1 bar
(DLG III) or 3 bar (DGM).
• Do not allow the flow to fall below the minimum rate of
20 l/h. Monitor the flow with the connected measuring
device/controller. If the sensor's signal is used for con-
trolling, switch off the controller or adjust it to constant
load when the flow rate falls below the lower limit.
• To guarantee the required flow conditions, the sensor
should be used only in the following in-line probe housing
models: DLG III A (914955), DLG III B (914956) or DGM
(25 mm module). Correct function cannot be guaranteed if
the sensor is used with other in-line probe housings.
• Avoid installations which allow air bubbles to enter the
sample water. Air bubbles clinging to the sensor
membrane can lower the resultant measured value and
cause incorrect dosing within the control system.
Slide the O-ring from below over the sensor shaft up to the clamping ring.
Insert the sensor into the DLG III.
Fix the sensor tightly by screwing in the plug.
Slide the O-ring from below over the sensor shaft up to the clamping ring.
Leave a flat washer inside the DGM.
Insert the sensor in the DGM and screw on the plug until the O-ring seals.
The clamping ring determines the installation depth of the sensor.
Assembly
33