Fig. 4
Electrical
connection of
the sensor
CTE 1-mA
CTE 1-DMT
®
ProMinent
Loosen the M12 threaded connector and guide the 2-wire cable into the
sensor. Leave an approx. 5 cm length of the 2-core measuring line in the
sensor.
Strip the two wires at their ends and connect them to the terminal block
by using the screwdriver provided: 1 = plus, 2 = minus (see figure 4).
Tighten up the M12 connector.
Insert the top part of the sensor right into the housing and turn the top
part clockwise as far as it will go.
Insert the 5-pin plug into the socket.
Installation
35