Technical Data - ProMinent CTE 1-mA Instrucciones De Servicio

Sensor de cloro total
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Technical Data

14 Technical Data
Measured
variable
Total chlorine:
Applications
Potable, industrial, process and cooling water,
brine and sea water for swimming pool and Jacuzzi treatment
Measurement
CTE 1-mA-0.5 ppm
range and
CTE 1-mA-2 ppm
standard slope
CTE 1-mA-5 ppm
CTE 1-mA-10 ppm
CTE 1-DMT-10 ppm
Solution
Corresponds to lower measurement range limit
Reaction time
T
90
pH range
6.5 – 9.5
pH-
dependency
Jump from pH 7 to pH 8: approx. -10 % for free chlorine
Conductivity
water sample
0.03 - 40 mS/cm
Temperature
range
5 - 45 °C, temperature compensation integrated, no quick changes in
temperature allowed
Pressure
sample water in DLG III in-line housing: max. 1 bar (free outlet)
sample water in DGM in-line housing:
Flow
sample water through DLG III and DGM in-line
probe housing:
Cross sensitivity
Oxidising reagents e.g. bromine, iodine, ozone, chlorine dioxide,
permanganates result in false positive results.
Reducing reagents like sulphides, sulphites, thiosulphates, and hydrazine
result in false negative results.
Service life
membrane cap
Typically 3 - 6 months, depending on water quality
Materials
Membrane cap:
Electrode shaft:
PVC
Supply voltage
16 - 24 V DC; min 35 mA at 16 V DC
Output signal
CTE 1-mA: 4-20 mA, temperature-corrected, un-calibrated and not
CTE 1-DMT: 0-1000 mV after connection to DULCOMETER
Enclosure rating
IP 65
Storage
temperature
between 5 °C and 50 °C
44
free chlorine (Cl
bound chlorine (chloramines)
organic bound chlorine (e.g. cyanuric acid derivates)
0.01 - 0.5 mg/l
0.02 - 2.0 mg/l
0.05 - 5.0 mg/l
0.01 - 1.0 mg/l
0.01 - 10 mg/l
approx. 60 s (as concentration rises and falls)
PPE
electrically isolated
(dissolved) HOCl, OCl - )
2
standard slope: 24 mA/ppm
standard slope: 6 mA/ppm
standard slope: 2.4 mA/ppm
standard slope: 1.2 mA/ppm
standard slope: 32.5 mV/ppm
max. 3 bar (no pressure spikes
allowed)
optimum:
30 l/h
minimum:
20 l/h
maximum: 100 l/h
®
DMT
®
ProMinent

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Cte 1-dmt

Tabla de contenido