Calibration of Lisa MB

Before carrying out a calibration, you must be sure that the sterilizer is vacuum-tight.
Run at least three vacuum tests: before connecting, after connecting and after disconnecting the sterilizer to/from the calibration
device. Fix any vacuum fault before proceeding.
Remove all trays from the chamber. The tray rack must remain placed inside the chamber.
Using the pressure sampling rubber tube, connect the steam pressure input of the
calibration device to the pneumatic test connection on the back side of the sterilizer.
Remove the metal cover of the sterilizer.
If required (see chip compatibility table on the right), replace the existing chips (see picture)
with the calibration chips provided with the device. Then, reinstall the cover.
Extract the chip carefully by pulling it outwards. Always use the tool
provided . Never touch the chips with your hands.
CHIP COMPATIBILITY TABLE
Board version
Chip A
1.0 or 1.1
CE.1.0.0
2.2
CE.2.0.0
2.2
CE.2.1.0
2.2
CE.3.0.0
Liquid cycle version only
Chip B
Chip A
ENGLISH
DEUTSCH
FRANÇAIS
ESPAÑOL
Chip B
DT3.0
DT3.1
MS1.03
USA – Canada Only
DT3.3L