Sterilizers that are not fitted with a serial communication port
The following sterilizers are not equipped with a serial communication port:
-
Lisa (VA131 2019)
-
Lara (RIL-1)
-
Lina (RIK-1
-
Lexa (MN-111)
-
Lyla (RIN-210)
These sterilizers need the Male serial port flat cable (see picture, included in the package) to be connected to the calibration device.
Access the sterilizer control board by removing the left side panel. On Lexa, it is necessary to remove both the side panels and the rear cover (See
"Repair and spare parts" section of the Service Book).
Plug the serial flat cable into the connector CON38 (see below pictures). Now the calibration device can be connected to the unit through the NULL-
MODEM female – female data cable.