On the calibration device, turn the model selector to the LISA MB position (see picture).
Connect the calibration device to the serial port of the sterilizer, located on the front side under the
service door, using the standard data cable fitted with the NULL-MODEM adapter.
For LISA MB, you must use the standard data cable (male – female) plus the
NULL-MODEM adapter (see picture), otherwise the communication between the
calibration device and the sterilizer won't take place
Connect the power supply to the calibration device.
Switch the sterilizer ON and insert a memory card into the Lisalog, if this option is installed.
Insert a memory card into the calibration device. This will help you in analyzing the calibration data, if needed.
From the start screen, access the list of the available cycles.
WARNING! It is possible you are requested to confirm or to key in the serial
number! If so, do it and proceed.
Choose the FACTORY CALIBRATION cycle and launch it.
Wait until the cycle is completed (time required: about 30 minutes). The sterilizer is now
calibrated.
Reinstall the original chip on the main board, if needed.
Check and, if necessary, set the date and time in the sterilizer.
Check the calibration is effective by running a 134°C cycle and a 121°C cycle.
Male – female data cable
NULL-
MODEM
adapter
12