Gima 33676 Manual Del Usuario página 182

Ocultar thumbs Ver también para 33676:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Recommended separation distances between portable and mobile RF communications
equipment and the Sleep apnea screen meter
Recommended separation distances between portable and mobile RF communications
The Sleep apnea screen meter is intended for use in the electromagnetic environment in which
radiated RF disturbances are controlled. The purchaser or the user of the Sleep apnea screen
meter can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the Sleep apnea
screen meter as recommended below, according to the maximum output power of the
communications equipment.
Rated power of
transmitter
(W)
0.01
0.1
1
Table 4
equipment and the Sleep apnea screen meter
Separation distance according to power of transmitter (m)
150 k Hz – 80 MHz
P
d=1.2
0.12
0.38
1.2
80 MHz – 800
MHz
P
d=1.2
0.12
0.38
1.2
79
800 MHz – 2.5 GHz
P
d=2.3
0.23
0.73
2.3

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Rs01

Tabla de contenido