Cleaning, disinfection and sterilization
The herein described reprocessing instructions are intended for the parts delivered within the set of the
OFA-Drill. Extension or accessory parts come with their own reprocessing instructions.
The following points in particular are important with regard to caring for the material:
● Perform cleaning, disinfection and sterilization after every treatment.
● Always autoclave the material in sterilization packaging.
● Make sure that sterilization packaging is no more than 80 % filled.
● Always autoclave the material at 135°C for at least 5 minutes.
● If sterilized material is not used immediately, the material packaging must be labeled with the
sterilization date.
● Nouvag AG recommends including a sterility indicator.
8.1
Control unit and pedal
The control unit and pedal do not come into contact with the patient.
Wipe the outside using micro-biologically tested surface disinfectant or a 70 % isopropyl alcohol solution.
The front plate of the control unit is sealed for this purpose and can be wiped clean.
OFA-Drill operating instructions, ref. 31595, V49/15