HighSurg 11
7.4
Adjusting the programs
Values for operation settings depend on the connected handpiece or contra angle as well as the task to be
performed.
7.4.1
Step 1: Selecting handpiece or transmission ratio
Depending on the handpiece attached to the motor, the corresponding transmission ratio has to be
adapted accordingly by using the "HP" or the "SH" key.
Press «HP» key to select the 1:1 Handpiece.
Press «SH» key to select Program 1 or 2 (Shaver programs).
Parameter of 1:1 Handpiece
Name of Handpiece with
transmission ratio
Handpiece, 1:1
Shaver functions table
Shaver function
Continuous Shaver
Oscillating Shaver
The oscillation period of the Shaver blade, attached at the Shaver Handpiece, is based upon the max. speed set. If
the oscillation period is set too low according to the selected speed, it is automatically adjusted to the nearest
possible value. Thus a complete revolution of the Shaver blade is guaranteed according to the selected max speed.
HighSurg 11, Operating Instructions, Ref. 31998, V14/15
Handpiece name
Speed
on Display
min. rpm
1 : 1
300
Speed
Display
min. rpm
Cont.Shaver
300
Osc.Shaver
300
Speed
Torque
max. rpm
min. Ncm
50,000
1
Oscillation
Speed
period
max. rpm
from ... seconds
6000
5000
0.20
EN
Torque
max. Ncm
6
Oscillation
period
to ... seconds
3.00
11