Table 3
For all NON-LIFE-SUPPORT medical-electrical (ME) DEVICES and ME SYSTEMS
Guidance and Manufacturer's Declaration – Electromagnetic immunity
The Veroval
upper arm blood pressure monitor is intended for use in the electromagnetic
®
environment specified below. The customer or the user of the Veroval
pressure monitor should assure that it is used in such an environment.
Immunity tests
Conducted RF distur-
bances IEC 61000-4-6
Radiated RF disturbances
IEC 61000-4-3
IEC 60601 test level
3 Vrms 150 kHz to
80 MHz
3 V/m 80 MHz to
2.5 GHz
96
®
Compliance
Electromagnetic
level
Environment –
Guidance
Portable and mobile RF
communications equip-
ment should be used
no closer to any part of
the Veroval
blood pressure monitor,
including its cables,
than the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended separa-
tion distance:
3 Vrms
3 V/m
800 MHz
2.5 GHz
upper arm blood
upper arm
®
80 MHz to
800 MHz to