TECNO-GAZ
Battery
specifications
Battery life
Indicator operation
Battery
specifications
Battery life
Accessory manual
5.1.1 Batteries for ORION40/40DS
Nominal voltage 12 V;
Capacity 12 Ah;
Maximum charge current 3 A;
Weight 3.75 kg;
Dimensions 151x98x94 (mm);
Operating range -20°C / 40°C.
The batteries have a life of about 4 hours.
The batteries are subject to a number of charge and discharge cycles.
- If the batteries are completely discharged the number of theoretical life
cycles is 250.
- If the batteries are discharged by 50% and subsequently recharged, the
number of theoretical life cycles is 550.
- If the batteries are discharged by 30% and subsequently recharged, the
number of theoretical life cycles is 1200.
The battery charge is indicated by a coloured LED.
If the light is green, this means the batteries are charged, if it is orange, the
batteries are being charged and if it is red this means the batteries are
down.
When the indicator is red, connect the lamp to the power mains.
5.1.2 Batteries for ORION60 ORION80
Nominal voltage 12 V;
Capacity 12 Ah;
Maximum charge current 6.8 A;
Weight 8.5 kg;
Dimensions 166x175x125 (mm);
Operating range -20°C / 40°C.
The batteries have a life of about 3 hours.
The batteries are subject to a number of charge and discharge cycles.
- If the batteries are completely discharged the number of theoretical life
cycles is 200.
- If the batteries are discharged by 50% and subsequently recharged, the
number of theoretical life cycles is 400.
- If the batteries are discharged by 30% and subsequently recharged, the
number of theoretical life cycles is 1200.
The system does not feature battery charge indicators.