4.6.5.2 Power up Zero
Power up zero can be disabled or enabled, and the range is
selectable from +/-2%, +/-10% or +/-20% of full capacity.
4.6.5.3 Pushbutton Zero
The pushbutton zero can be set to +/-2%, +/- 10% or +/- 20%
of full capacity.
Tare
4.7.1 Automatic Tare
When auto tare is enabled, the display changes to a zero net weight
indication after the weight exceeds the threshold value. Auto tare
includes:
• Auto tare threshold Weight: When weight on the scale platform
exceeds the tare threshold value, the terminal automatically tares.
• Auto tare reset Threshold Weight: The reset threshold weight must
be less than the tare threshold weight. When the weight on the scale
platform falls below the reset threshold value, such as when a load
has been removed, the terminal automatically resets the auto tare
trigger.
4.7.2 Clearing Tare
Tare values can be cleared manually or automatically.
4.7.2.1 Manual Clear
Press
on the scale will not impact a manual clear.
4.7.2.2 Auto Clear Tare
The IND236 can be configured to clear tare automatically when
the weight returns to zero point. Once the tare is cleared, the
display returns to the gross weighing mode.
4.7.3 Keyboard Tare
Pushbutton tare can be configured enabled or disabled. When it is
disabled, pressing
Terminal Configuration.
If pushbutton tare is enabled, pressing the pushbutton
semi-automatic tare. The IND236 will perform a tare process. If the
process is successful, the display changes to the zero net weight statues
and the current weight will be stored as the tare value. The net symbol
will be on the display.
to clear tare when the IND236 is in net mode. Motion
has no effect. please refer to F1.5.1 in Chapter of
initiates a