4.9.1.3 Viewing Configured Values
When in the check-weigh application, press and hold the Target
function key to view the current active target, tol- and tol+ values.
The information will display for one second, then the target value
will appear together with
repeatedly to cycle through the three check-weigh values - OK,
low tolerance and high tolerance. Press
4.9.2 Counting
The IND236 counting application provides a simple counting sequence
to determine a count value.
4.9.2.1 Operation procedure
1. Press the counting function key
displayed, then "PCS 05", as shown in Figure 5-4.
2. The reference number can be adjusted using
(increase) keys. Selections are 5, 10, 20, 50 and 100.
3. Confirm the reference number by pressing
display (Figure 5-5) while the scale detects stable weight.
Make sure that the number of samples on the scale matches
the selected reference size. If stable weight is detected
before the process times out, the counting application starts;
otherwise the screen remains in the previous weighing state.
4. Load the parts to be counted. The count will be displayed
(Figure 5-6).
5. When counting is complete, press
application application and return to the basic weighing interface.
. Press the "Target" key
Figure 5-4: Reference Number
Figure 5-5: Detecting Stable Weight
Figure 5-6: Piece Count Display
to exit.
. "SAMPLE" will be
(decrease) and
. "------" will
to exit the counting
23