with the normal AccelStim device operation. To restore normal operation, press
the reset button on the left side of the AccelStim device with a pointed object,
and turn on the device. Be sure to remove the source of disturbance before
continuing the treatment if closer than 30 cm (12 inches).
Battery and Charging Safety
To ensure that the device is functioning properly, the AccelStim device constantly
monitors battery voltage level in treatment mode and displays on the upper right
corner of the screen. When the battery level decreases to low battery level, the plug
symbol and the battery level shown in red indicate the need to connect the external
power supply to recharge the battery. In the low battery condition, treatment delivery
can continue.
If not recharged, the battery level decreases to an empty battery level, indicated by a
flashing empty battery symbol and three repeated short beeps. The AccelStim device
will automatically stop treatment. To continue treatment, connect the power supply
and press the function button.
Recharging the AccelStim device Battery
To recharge the battery within the AccelStim device, use the following steps:
NOTE: When the power supply is connected to the AccelStim device while the device is
in treatment mode, the charging process starts and a flash symbol appears next to the
battery status symbol.
WARNING: In case of a faulty power supply, the device is not powered on or charging is
not started, contact Patient Services at 1-800-535-4492 for assistance.
17
1. Open the charging port cover on the generator.
2. Plug the power supply DC plug into the charging port located on the
generator.
3. Plug the power supply into an AC wall outlet.
4. When the power is connected, the generator emits a short "beep" and
the battery charging process started. During the recharging battery
process, the battery charge status indicator moves from one level to
the next until process is finished. Please refer to Visual and Audio Battery
Indicators on page 20 for more information.
5. When the charging process is completed the full battery symbol is
displayed on the screen. Remove the power supply and insert the rubber
piece back into the charging port.