Interrupted Treatment / Calendar Indicators
The AccelStim device will disrupt treatment if alarm conditions occur. You will be
notified by visual sounds and/or audio messages described in this section. When
possible, restore the normal condition and restart treatment by pressing the function
button.
Display Message
21
Audio Signals
Transducer Not Connected. Check the
Three short
connection of the transducer to the AccelStim
beeps every 3
device and press the function button to start
seconds.
the treatment.
Treatment Not Allowed. The AccelStim device
allows a maximum of two treatments per
day. Only complete two treatments per day if
instructed by your physician. If two treatments
Three short
have been completed, no other treatments are
beeps every 3
allowed. The AccelStim device will switch off
seconds.
automatically after 30 seconds.
*The second treatment must be completed by
midnight on the current day.
Expired Device Life. The AccelStim device
Three short
provides daily treatment for up to 365 days
beeps every 3
from date of first use. The AccelStim device
seconds.
switches off automatically after 30 seconds.
Contact Patient Service at 1-800-535-4492.
Fault Detected. If the AccelStim device detects
an anomaly in the transducer or device
operation, treatment is stopped. Check the
Three short
presence of gel on the transducer which must
beeps every 3
form a 1-2 mm thick layer to the transducer.
seconds.
Then restart the treatment by pressing the
function button. If the message remains after
checking the gel, turn the device off and contact
Patient Services at 1-800-535-4492.
Over Current Fault Detected in Calendar
or stand-by mode. If over current fault
detected, the AccelStim device will switch off
Three short
automatically after 5 seconds. Contact Patient
beeps every 3
Services.
seconds.
NOTE: If detected over current condition when
export treatment data to USB portable drive,
please try again while device is charging or after
the device battery is fully charged.
Problem and Solution