TROUBLESHOOTING THE ACCELSTIM DEVICE
The AccelStim device will disrupt operation if alarm conditions occur. You will be
notified by visual sounds and/or audio messages described in this section. If further
technical assistance is required, please contact Patient Services at 1-800-535-4492.
Visual and Audio Battery Indicators
Battery charge status is shown on the upper right corner of the display.
When the battery is fully charged, the AccelStim device can deliver up to
five treatments.
The battery symbol is characterized by several notches, which decrease
as the battery runs down.
The battery level will decrease gradually until it reaches the low battery
level. The red indicator and the plug symbol will display indicating the
need to connect the power supply to recharge the battery.
When the battery level is low, connect to the power supply using the
instructions on page 16.
When the power supply is connected to the AccelStim device, the
charging process starts and a flash symbol appears next to the battery
status symbol.
WARNING: In case of a faulty power supply, the flash symbol does not
appear. Contact Patient Services for assistance.
Charge & Treat Mode:
During the battery charging process, the battery charge status indicator
moves from one level to the next until the process is finished. When the
charging process is completed the full battery symbol is displayed.
Charge Only Mode:
It is possible to recharge the battery when the AccelStim device
is turned off.
If a faulty battery is detected when the device is turned on and the
charging process is not possible, a battery symbol with the warning
symbol (triangle/exclamation point) will appear on the screen. Please
contact Patient Services at 1-800-535-4492 for assistance.
In case of faulty battery, it is possible to complete a treatment by
connecting the external power supply to the AccelStim device. After a
few seconds, the AccelStim device will automatically start in treatment
mode. At the end of the treatment, it is necessary to disconnect the
power supply from the AccelStim device to power down the device.
In case of faulty battery, it is not possible to switch the AccelStim device
to calendar mode. Battery charge status is shown on the upper right
corner of the display. The X symbol on the battery indicates that the
battery is faulty and is not charging.
20