STORAGE AND OPERATING ENVIRONMENTS
When moving the AccelStim device from very cold or very hot storage areas (like your
car), wait at least an hour to use or charge the device. The device requires time to
return to a safe operating temperature
Environmental Operating Conditions (Lower / Upper Limits)
Ambient temperature:
Relative humidity:
Atmospheric pressure:
Environmental Conditions for Transport and Storage
The system can be transported and stored at the following environmental conditions
without risk of any deterioration. NOTE: After removing the device from its protective
packaging, the environmental operating conditions are applicable for transport and
storage between uses.
Ambient temperature
Relative humidity
Atmospheric pressure
The AccelStim device is designed for a storage life of 12 months and one year of
usage.
THE ACCELSTIM DEVICE CLASSIFICATIONS
Product Family Name: Orthofix AccelStim Device.
Equipment is internally powered or may be used with provided external power
supply.
This device generates a low-intensity pulsed ultrasound with a frequency of
1.5 MHz ± 5%. This ultrasound is an acoustic vibration with frequency above the
human auditory level, thus the device is silent.
Storage life for equipment: 12 months.
Mode of operation: intermittent operation.
This device is nonsterile. It does not require sterilization.
Equipment not suitable for use in the presence of a flammable anesthetic
mixture with air or nitrous oxide.
The power supply is considered double insulated with Class II construction.
23
10/35°C
15%/93% (non-condensing)
700/1060hPa
Transport
-20/+60°C
10%/90%
(non-condensing)
500/1060hPa
Storage
5/30°C
15%/93%
(non-condensing)
700/1060hPa