GENERAL INFORMATION
IEC/EN 60601-1 classification: Device of Class II - Applied part of type BF
Expected lifetime: 2 years.
Generator plastic case: IP22 degree of protection.
Ultrasound transducer case: IPX7 degree of protection.
Battery Internally powered equipment: Li-Ion rechargeable battery - 3,7VDC -
1100mAh - 4,07Wh.
The AccelStim device is isolated from the supply mains by means of a Class II external
power supply.
The first digit (IP2x) expresses the degree of protection against the entry of solid objects. Degree of
protection 2 means that the device is protected against the entry of solid objects larger than 12 mm Ø
(e.g. a finger).
The second digit (IPx2) expresses the degree of protection against the ingress of liquids. The degree of
protection 2 means that the device is protected against dripping water at an angle within ± 15°.
External power supply used as battery recharger
Model*
Brand
Input power
Mains voltage
Max. input current
Output voltage
Max output current
Short-circuit protection
Insulation class
OPERATING SPECIFICATION
Ultrasound frequency:
Pulse width:
Repetition rate:
Duty factor:
Effective radiating area (ERA):
Temporal average power:
Effective intensity ISATA:
Beam non-uniformity ratio (BNR):
Beam type:
ME10A0503B01
SL POWER ELECTRONICS
100 - 240 VAC
50 – 60 Hz
0,3 A
5 VDC
2,0 A
Internal self-resetting
protection - Continuous
II
1.5 MHz ± 5%
200 μsec ± 10%
1 KHz ± 10%
20%
3.5 cm
110 mW ± 10%
30 mW/cm
3.8 ± 30%
Collimated
*Orthofix reserves the right
to provide different models
of power supply, tested and
approved for the system
according to the standard
EN60601-1; use only the
power supply provided.
2
± 30%
2
24