9.4.1
Realizar un ajuste de cero
Visualización
Calibration
Zero Adjust
Zero Adjust
Span Test
Span Test
back
Password
Password 1234
Password 1234
back
Zero Adjust
Are you sure to
Are you sure to
start adjust
start adjust
procedure ?
procedure ?
back
Zero Adjust
Caution operation
Caution operation
temperatur not
temperatur not
valid
T: 61.5°C
T: 61.5°C
back
Zero Adjust
Please Wait !
Please Wait !
***********
Zero Adjust
Please wait
Please wait
Amplifer Values
Amplifer Values
Amp1: 0
Amp1: 0
Amp2: 6
Amp2: 6
*************
back
Zero Adjust
C1
C1
: +0,0...
: +0,0...
C2 -var
C2 -var
: +0,0...
: +0,0...
C3
C3
: +0,0...
: +0,0...
back
8014820/1F7Z/V4-0/2022-06 | SICK
Sujeto a cambio sin previo aviso
Acción
▸
Pulse la tecla "CAL"
▸
Active la selección "Zero Adjust" con la
tecla "Enter"
▸
Introduzca la contraseña "1234"
select
▸
Introduzca "Enter"
▸
Cancele con la tecla "flecha a la
izquierda" (back)
Start: Enter
Start: Enter
Start: Enter
Start: Enter
▸
Introduzca "Enter"
(a T=60 °C +- 0,5 °C)
▸
Cancele con la tecla "flecha a la
izquierda" (back)
▸
Introduzca "Enter"
Save: Enter
Save: Enter
Nota
Sólo hay una interrogación en caso de
una advertencia pendiente (p. ej.
temperatura del dispositivo)
Espere hasta que se haya alcanzado la
temperatura del dispositivo.
Sólo está indicado un mensaje si la
temperatura todavía no se ha estabili-
zado
Durante el proceso de ajuste no es
posible realizar entradas en el disposi-
tivo
Durante el proceso de ajuste no es
posible realizar entradas en el disposi-
tivo
Se guardan los datos
I N S T R U C C I O N E S D E S E R V I C I O | GM901
9
MENÚS
69