Publicidad

Idiomas disponibles

Idiomas disponibles

4.
Select Set automatically ON and enter the settings for your time server.
5.
Choose 12- or 24-hours clock and select NEXT.
NTP provides your trophon2 device with Co-ordinated Universal Time (UTC). Your device will use this
to calculate the time in the local time zone that has been set. The time setting will periodically
synchronize with the UTC. The device will update the time when DST starts and ends according to the
timezone set.
Printer label
Select the number of printer labels; up to 4 records can be printed.
Daily timers
Configure a preset daily wake and sleep time.
Brightness and Alarm
Configure the brightness of the trophon2 touch screen and the volume of the trophon2 device alarm.
Network
Network access is required to allow for automatic time setting.
Consult your facility's IT administrator or network parameters expert to complete this setting.
The device supports the TCP/IP protocol. You can set a static IP address, or you can use the Dynamic
Host Configuration Protocol (DHCP), which means that the router will assign a valid IP address. Select:
Menu
Settings
Network
Then choose DHCP or Static IP. If you use a static IP address, the screen will request network
parameters. Your network administrator can provide an IP address that is not in use and will advise
appropriate values for the subnet mask and the default gateway.
For details concerning trophon AcuTrace
®
B2.7 AcuTrace
AcuTrace is an automated HLD traceability system that uses Radio Frequency Identification (RFID).
AcuTrace-enabled consumables and accessories contain RFID chips that store information and can be
read by the built-in AcuTrace reader on the trophon2 device.
Each medical instrument (ultrasound probe) is allocated a trophon AcuTrace Medical Instrument Tag
and individual trophon2 operators are assigned a trophon AcuTrace Operator Card. This is important to
meet compliance documentation requirements in many regions.
The reprogramming of Medical Instrument Tags and Operator Cards is not recommended, as
this may impact the disinfection traceability system.
®
PLUS, see B3.2 Network Parameters Setup
18
L03192 3.0 06/2022