C3.4 Closing the Chamber Door
The chamber door utilizes a two stage closure mechanism. Carefully close the chamber door to
the first click and do not force it shut. The door will automatically close further to seal and lock at
the start of the HLD cycle.
If the door is not properly closed, a screen message will prompt you to close the chamber door.
DO NOT force open or manually unlock the chamber door during a HLD cycle.
C3.5 Disinfecting the Probe
Once the door is closed, the trophon2 device will request confirmation that the probe is clean and dry
prior to HLD. The door will lock upon confirmation.
7.
Scan the trophon AcuTrace Operator Card when prompted (if AcuTrace is enabled) and follow
the onscreen instructions.
The screen will ask:
Is the probe clean and dry?
8.
If the probe has NOT been cleaned and dried, select No. Remove the probe, clean and dry the
probe in accordance with C3.1 Preparing the Probe and repeat from C3.3 Positioning the Probe.
If the probe is clean proceed with the step below.
9.
Once the probe is cleaned and dried and the probe has been correctly inserted into the trophon2
device, press the cycle start symbol
If you wish to unlock the device and remove the probe, select Cancel and follow the onscreen
instructions.
10.
The progress of the disinfection cycle is indicated on the screen.
11.
The HLD cycle will take 7 minutes to complete.
In the unlikely event that mist is escaping from the chamber, avoid direct contact with the
mist and remain at distance from the trophon2 device until completion of the operating
cycle and the mist is no longer visible. Contact your customer service representative.
Refer to Part D – Troubleshooting for further information.
a
Figure 15
on the touch screen to initiate the cycle.