5.
Ensure that the probe is correctly positioned in the chamber. The probe must not make contact
with the chamber wall and must be positioned at or above the embossed line at the bottom of the
chamber. See Figure 11.
6.
If the probe is disconnected from its connector end, the external portion of the cable and connector
can be secured by placing the connector carefully inside the cable tray.
NOTE: Incorrect positioning of the probe may result in:
Unsuccessful HLD cycles.
Residual disinfectant remaining on the probe's surface. This may lead to temporary bleaching or
irritation of the skin, or both, if gloves are not worn.
Uncertain probe compatibility as stated in the trophon EPR and trophon2 Compatibile Ultrasound
Probe List.
It is customer's responsibility to ensure the probe is correctly positioned in the chamber. Probe
compatibility with the trophon2 device and successful HLD cannot be guaranteed if probe is incorrectly
positioned in the chamber.
NOTE: Curved probes must be correctly inserted in the trophon2 device. See Figure 13.
Figure 11
24
L03192 3.0 06/2022
Figure 12