Descargar Imprimir esta página

Medtronic GUARDIAN 4 Manual Del Usuario página 17

Ocultar thumbs Ver también para GUARDIAN 4:

Publicidad

Idiomas disponibles

Idiomas disponibles

2. Remove the adhesive tab from the top of the transmitter.
3. Hold the transmitter as shown, and pinch the flexible side arms of the sensor between the
thumb and forefinger.
4. Gently pull the transmitter away from the sensor.
Removing the sensor
Always refer to the Guardian 4 Sensor User Guide for instructions on how to remove the sensor.
Reconnecting the transmitter to a sensor that is already inserted
The transmitter can be reconnected to the sensor currently in use. Simply connect the transmitter to
the sensor that is already inserted. Select Reconnect Sensor when the display device detects the
transmitter. It may take a few seconds to establish a connection when reconnecting a sensor. Reattach
the adhesive tab of the sensor to the transmitter and reapply any required tape. When reconnected,
the sensor goes through another warm-up period.
Tester
The tester is used to test the transmitter to make sure it is working. The tester is also used as a required
component to create a waterproof seal when cleaning the transmitter. Properly connecting the tester
to the transmitter ensures that fluids do not come in contact with the connector pins inside the
transmitter. Fluids can cause connector pins to corrode and affect the performance of the transmitter.
Do not twist the tester while attached to the transmitter. This will damage the transmitter.
The tester can be used for one year. If the tester is used for more than one year, the connector pins inside
the transmitter can be damaged, because the tester cannot continue to provide a waterproof seal. For
instructions on how to check the connector pins, see Inspecting the transmitter connector pins,
page 18.
17

Publicidad

loading