Table 1. Troubleshooting issues (continued)
Problem
When connected to the
sensor, the green light
on the transmitter does
not flash.
When connected to the
tester, the green light on
the transmitter does not
flash.
The transmitter battery
does not last for seven
days.
Likely Cause(s)
The transmitter is not ful‐
ly connected.
The transmitter is low on
power.
The sensor is not proper‐
ly inserted into the body.
The transmitter is not ful‐
ly connected.
The transmitter is low on
power.
The transmitter is not
fully charged when con‐
nected to the sensor.
The transmitter and
display device frequent‐
Resolution
call 24-Hour Technical Support. It may be
time to replace the transmitter.
1. Disconnect the transmitter from the sen‐
sor.
2. Wait for five seconds and reconnect
them. If the green light still does not
flash, proceed to step 3.
3. Fully charge the transmitter and connect
it to the tester. If the green light still does
not flash, see troubleshooting on "When
connected to the tester, the green light
on the transmitter does not flash. " If the
green light flashes, proceed to step 4.
4. Disconnect the transmitter from the
tester, wait at least five seconds, and
connect the transmitter to the sensor.
If the green light still does not flash,
proceed to step 5.
5. The sensor may not be properly inserted
into the body. Remove the sensor from
the body and insert a new sensor.
1. Check the connection between the
transmitter and the tester. If the green
light still does not flash, proceed to step
2.
2. Fully charge the transmitter.
3. Test the transmitter with the tester again.
If you still do not see the green light flash,
call 24-Hour Technical Support. It may be
time to replace the transmitter.
1. Fully charge the transmitter before con‐
necting it to the sensor. If the transmitter
battery still does not last for the duration
of one sensor use, proceed to step 2.
2. Move away from any equipment that can
cause RF interference. For more informa‐
26