HBP-1320-E main.book Page 34 Tuesday, May 8, 2018 5:16 PM
Manufacturer's Declaration
The HBP-1320 is intended for use in the electromagnetic environment specified below.
The customer or the user of the HBP-1320 should assure that it is used in such an environment.
Electromagnetic Emissions:
(EN60601-1-2)
Emission Test
RF emission
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
IEC 61000-3-3
Electromagnetic Immunity:
(EN60601-1-2)
Immunity test
Electrostatic discharge
±8 kV contact
(ESD)
±2 kV, ±4 kV,
IEC 61000-4-2
±8 kV, ±15 kV air
Electric fast transient/
±2 kV for
burst
power supply lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
±1 kV
IEC 61000-4-5
normal mode
±2 kV
common mode
Voltage dips, short
0 % U
interruptions and
At 0˚, 45˚, 90˚,
voltage variations on
135˚, 180˚, 225˚,
power supply input lines
270˚ and 315˚
IEC 61000-4-11
0 % U
70 % U
25/30 cycles
single phase: at 0˚
0 % U
250/300 cycles
Power frequency
30 A/m
(50/60 Hz) magnetic
50 Hz or 60 Hz
field
IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level.
T
34
Compliance
Group 1
The HBP-1320 uses RF energy only for internal functions.
Therefore, this RF emission is extremely weak and there is
little chance of it creating any kind of interference
whatsoever with nearby electronic equipment.
Class B
The HBP-1320 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low voltage power supply network
Class A
that supplies buildings used for domestic purposes.
Complies
EN60601-1-2
Compliance
test level
level
±8 kV contact
±2 kV, ±4 kV,
±8 kV, ±15 kV air
±2 kV for
power supply lines
±1 kV for input/
output lines
±1 kV
normal mode
±2 kV
common mode
; 0.5 cycle
0 % U
; 0.5 cycle
T
T
At 0˚, 45˚, 90˚,
135˚, 180˚, 225˚,
270˚ and 315˚
; 1 cycle
0 % U
; 1 cycle
T
T
;
70 % U
;
T
T
25/30 cycles
single phase: at 0˚
;
0 % U
;
T
T
250/300 cycles
30 A/m
50 Hz or 60 Hz
Electromagnetic Environment
Electromagnetic
environment - guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
HBP-1320 requires continued
operation during power mains
interruptions, it is recommended that
the HBP-1320 be powered from an
uninterruptible power supply or a
battery.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.